HP CodeMaster XL Specifications

Dimensions           

7.9”H x 11.8”W x 15.6”L/20 x 30 x 39.7 cm)

Weight                   

24 lbs./10.9 kg

Battery                   

Type: Rechargeable, sealed lead-acid battery.  4 Ah, 12 V nominal.  Charge time: 2 hours to 90% of full capacity.  18 hours to 100% capacity.  Repeated charging to less than 100% will reduce the useful life of the battery.  Capacity: 2.5 hours monitoring or 50 full-energy discharges or 1 hour monitoring and recording.  AC input: 100 to 230 VAC ±15%, 50/60 Hz. Battery indicators: illuminated LED indicates battery is charging.  Low Battery message appears on the monitor

Defibrillator          

Waveform: damped sinusoidal.  Output energy (delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, and 360 joules.  Charge control: push-button on apex paddle and on front panel.  Charge time: less than 5 seconds to 360 joules with a functional battery installed.  Charge-done indicator: display indicates available energy, charge-done light on apex paddle, charge done tone.  Synchronizer: SYNC message appears on the monitor and is annoted periodically on the recorder while in synchronous mode.  An audible beep sounds with each detected R-wave, while a marker on the monitor and sync mark on the recorder strip indicate the discharge point.  Paddles: standard paddles are anterior/anterior.  Adult electrodes (83 cm sq) slide off to expose pediatric electrodes (21 cm sq).  Paddle cord is 10 feet (3m).  Full range of internal paddles available.

Monitor                  

Input: ECG activity may be viewed through the paddles or patient cable.  Lead fault: LEADS OFF message and dashed baseline appear on the monitor if a lead becomes disconnected.  Common mode rejection: greater than 100 dB measured as per AAMI standards for cardiac monitors.  Heart rate display: digital read out on monitor from 20 to 280 bpm.

Recorder                

Annotations: time, date, heart rate, event markers, ECG mode, and selected energy.  Recorder mode: automatically documents events and ECG during defibrillation episodes.  The recorder can be configured to run in either real time or with a 6-second delay.  Frequency response: 0.5 to 40 Hz monitor bandwidth.  0.05 to 150 Hz diagnostic bandwidth is selectable

HP CodeMaster XL defibrillators