Marquette MAX 1 Specifications

Display

Size: 30.48 cm (48 in) diagonal.  Type: high-resolution, raster written.  Frequency response: 0.01 to 40 Hz, 0.01 to 20 Hz, user-selectable.  Number of traces: 3 rhythm leads.  Sweep speed:  25 mm/s.  Sweep speed accuracy: ±2%.  ECG rhythm: 7.5 s/lead.  Leads: user-selectable.  Recall reports: 10 s of 6 user-selectable leads.  Displayed data: heart rate, target heart rate, blood pressure, total exercise time clock, stage clock, 4 current median complexes, treadmill speed and grade, ventricular ectopic/minute counter, 3 channels of rhythm, protocol and stage of exercise, warning messages and prompts.

System

Digital sampling rate: 250 Hz.  A/D conversion: 12 bits.  Chassis leakage current:  <100 mA (meets AAMI standards SCL 1278).  Patient leakage current: <10 mA.  Common mode rejection: 130 dB, minimum, up to 180 dB with digital processing.  AC elimination:  50/60 Hz digital filters.  Electrode offset tolerance: ±320 mVDC.  Internal maintenance test: automatic self-tests.  Diskette drive: 8.89 cm (3.5 in), 2 MB, double sided, quad-density, and MS-DOS format compatible.

Processing

ST level and slope determination: up to 15 leads.  Medians: incremental updating with elimination of artifact and aberrance in all leads.  ST level measurement: manually selectable post J point.  Heart rate: 16-beat sliding average updated every beat; minimum 35 beats/min, maximum 300 beats/min.  QRS detection: band-pass filtered composite of three user-selectable leads.  Arrhythmia: automatic detection and documentation.

Input

Channels: simultaneous acquisition of up to 15 channels.  Impendence: 100 M1/2, defibrillator-protected.  Dynamic range: ±320 mVDC.  Electrode offset tolerance: ±320 mVDC.

Output

Standard outputs: one analog output, one waveform output, one TTL output with variable pulse width, polarity and delay, QRS beep.

Writer

Method: thermal dot array (horizontal 1000 dots/in at 25 mm/s, vertical 200 dots/in).  Channels: 6, user selectable.  Frequency response: 0.01 to 100 Hz, 0.01 to 40 Hz, 0.01 to 20 Hz, user-selectable.  Sensitivity:  5, 10, and 20 mm/mV.  Paper speed: 5, 12.5, 25, and 50 mm/s, user-selectable.  Paper style: thermal, perforated, 21.59 cm x 27.94 cm x (8.5” x 11”), Z fold, 300 sheets/pack 

Acquisition Module

Analog Sampling rate:  2000 Hz.  Microprocessor-controlled:  10 IEC and AHA color-coded lead wires standard, 14 lead wires optional; (additional lead wires user-definable)

Keypads

Control panel: contains manual treadmill, test/procedures, and report controls.  Alphanumeric keyboard:  with TRIM KNOB for blood pressure, patient demographics, and impressions entry.

Reports

Classical 12-lead: note: reports can be transferred to the NUSE network by diskette.  Incrementally averaged medians:  with real-time rhythm and ST level; user-selectable in two formats.  3- or 6-channel rhythm strips: available for real-time, recall or arrhythmia documentation (leads user-selectable).  Tabular presentation of final report data including: phase, stage, speed, grade, heart rate, blood pressure, rate-pressure product, ectopic, maximum ST depression, and physician’s impressions.  Sample cardiac cycles: presented at user-defined intervals.  Trend graph: heart rate, ST levels, ST slopes, blood pressure, PVCs/m.

Power

Requirements: 90 to 135 VAC, 180 to 270 VAC, 50/60 Hz.  Power consumption: 225 W maximum.

Physical

Height: 154.94 cm (61 in).  Width: 62.23 cm (24.5 in).  Depth: 71.12 cm (28 in).  Weight: 54.48 kg (120 lbs)

Marquette MAX 1 cardiac stress tests