Dimension
13.3cm x 40.6 cm x 37 cm; 5.3” x 16” x 14.6” (h x w x d)
Weight
10 kg; 22 lbs
Battery
3 NiCad batteries, 12V, 1.0 amp hours each. Anew Battery Pack registering at least 100% capacity on the Battery Support System will provide at a minimum: 45 minutes of monitoring or 30 minutes of pacing or 25 discharges at 360 joules per Battery Pack
ECG Lead Selection
Paddles I, II, III
Input
Isolated ECG via Quick-combo® defibrillator paddles, 3-lead patient cable, or Quick-combo® pacing/defibrillator/ECG electrodes, or ECG electrodes
Cardioscope Display
Size
72.5 mm (2.85 in) x 43.5 mm (1.7 in), non-fade
Defibrillator Waveform
5msec monophasic pulse (Edmark) per AAMI spec
Energy Select
0, 5, 10, 20, 50, 100, 200, 300, 360 joules
Paddle Controls
Charge (with indicator light).; Record (activates strip chart recorder); Energy select (rotating dial 0-360J); Discharge (dual energy discharge buttons, one on each paddle)
Charge Time
Charges to 360 joules in less than 12 seconds above 0°C.
Paddle Area
82 cm²
Synchronizer
Delivers energy discharge within 20msec of sync marker on cardioscope
Strip Charge Recorder
Paper Size
50 mm x 30 m
Paper Speed
25 m /sec
Annotation
Includes time, date. lead, gain, heart rate, defibrillator and/or pacing parameters.
Critical Event Record
Digitally stored record of critical ECG and device parameters.