Zoll PD 1600P Specifications

General                                  

Size: 10.7 cm high x 33.5 cm wide x 31 cm deep (4.2 x 13.2 x 12.2 in).  Weight: 6.3 kg (14 lbs) with multi-function cable.  Power: sealed lead acid battery- 2.5V/cell-wired in series.  Patient safety: all patient connections are fully isolated.

Defibrillator                          

Waveform: damped sinusoid.  Charge time: less than 10 seconds.  Depleted battery packs will result in a longer defibrillator charge time.

Semi-automatic Defibrillator             

Advisory: evaluates electrodes connection and patient ECG to determine if defibrillation is required.  Shockable rhythms: ventricular fibrillation with amplitude > 150 µV and wide QRS complex ventricular tachycardia with rates greater than 150 bpm.  Output energy delivered: automatic preset shock 1, 2, 3 energy levels - user configurable.  Selectable at 50, 100, 150, 200, 300, 360 joules.  Voice prompts: stand clear, press shock, noshock advised/check pulse, analysis halted, check patient/press analyze, check electrodes, memory card full, check memory card.  Visual prompts: analyze, analyzing ECG, shock advised, joules selected, charging, press shock, number of shocks delivered, no shock advised, noisy ECG, check patient.

Manual Defibrillator            

Manual defibrillator access: keyswitch – removable or permanent, user configurable.  Output energy delivered: selectable at 2, 3, 5, 7, 10, 20, 30, 50, 100, 150, 200, 300, 360 joules.  Energy selection and charge control are on front panel.  Synchronized mode: synchronized defibrillator pulse to patient’s R-wave.  “SYNC” message displayed on monitor.  Marker on monitor and recorder paper identifies R-wave.  Smart alarms: beeper/voice prompts indicate ventricular fibrillation with amplitude > 150 µV and wide QRS complex ventricular tachycardia with rates greater than 150 bpm

Pacemaker                             

Type: VVI demand; asynchronous (fixed rate) when used without ECG leads.  Pulse type: rectilinear, constant current.  Pulse duration: 40 msec.  Pulse amplitude: variable to 140 mA.  Pacing rate: variable from 30 to 180 bpm.  Output protection: fully defibrillator-protected and isolated.

Batteries                                

Type: rechargeable, sealed lead-acid.  Charger: single battery charger, battery support system or powercharge.  Recharge time: 16 hour recharge time with single battery charger.  Low battery indicator: message display on monitor with audible tone.  Approximately 20 minutes of monitoring remain when message is displayed.  Operating time: with fully charged battery – 35 defibrillator charges to maximum energy (360J) or 2 hours of continuous monitoring.